dc.description.abstract | En la presente investigación se hizo un estudio de diversos filtros digitales que
puedan cumplir con la tarea de filtrar, en tiempo real, y usando una tarjeta de
adquisición de datos (TAD), señales PPG obtenidas para calcular la hemoglobina
en la sangre de una persona. Es por esto que, la tarea de filtrar estas señales
fotopletismográficas (PPG), es crucial, ya que un mal filtrado puede terminar en un
mal cálculo de hemoglobina. El primer paso fue estudiar el estado del arte alrededor
del filtrado de señales PPG y así determinar cuáles pueden ser las opciones para
hacer el filtrado. Posteriormente, se obtuvieron señales PPG sin filtrar de pacientes
para su estudio, lo que permitió determinar los parámetros para los filtros elegidos.
Luego de ello se determinaron las ecuaciones y los algoritmos para poder hacer la
comparación necesaria para la determinación del filtro. Una vez determinadas las
ecuaciones y algoritmos, se procedió a hacer su implementación en PyCharm,
usando el lenguaje de programación Python, lo que permitió determinar los
indicadores para la comparación de los filtros y la determinación del más eficiente,
es decir, que optimice los recursos computacionales disponibles sin consumo
excesivo. Una vez realizada la comparación, se determinó, según las necesidades
del proyecto, cuál es el filtro que cumplía los requerimientos, lo que resultó en el
filtro Butterworth de orden 6. Con la determinación del filtro, se procedió a
desarrollarlo en lenguaje C para luego implementarse en el microcontrolador del
proyecto, validando que el filtro, funciona según los requerimientos previamente
establecidos. | es_PE |
dc.description.abstract | In the present investigation a study of many digital filters was made that may
accomplish the task of filtering, in real time, and using a data acquisition board
(DAQ), PPG signals obtained to calculate hemoglobin in a person’s blood. This is
why, the task of filtering these PPG signals, is crucial, because, a bad filtering, may
result in a bad hemoglobin calculation. The first step was studying the state of art
surrounding photoplethysmographic signal (PPG) filtering, that way, determining
which options may do the filtering task. After that, unfiltered PPG signals were
obtained from patients, for its study, that way, the parameters needed, could be
determined for the study of the chosen filters. After that, the equations and
algorithms needed were determined for making the comparison for the filter
determination. Once the equations and algorithms needed were determined, the
implementation in PyCharm was done, using Python programming language, which
allowed us to determine the indicators for the filter’s comparison and the
determination of the most efficient one, that it optimizes the available computational
resources without excessive consumption. When the comparative table was done,
it was determined that, following the project needs, the most adequate filter, turned
to be order 6 Butterworth filter. With this result, it was developed the filter in C
language so it could be implemented in the microprocessor of the project, validating
that this filter, works according to the previously established requirements. | en_US |
dc.description.uri | Tesis | es_PE |
dc.format | application/pdf | es_PE |
dc.language.iso | spa | es_PE |
dc.publisher | Universidad Privada Antenor Orrego | es_PE |
dc.relation.ispartofseries | T_ELECT_879 | |
dc.rights | info:eu-repo/semantics/openAccess | es_PE |
dc.rights.uri | https://creativecommons.org/licenses/by/4.0/ | es_PE |
dc.source | Universidad Privada Antenor Orrego | es_PE |
dc.source | Repositorio Institucional - UPAO | es_PE |
dc.subject | Filtros | es_PE |
dc.subject | Señales | es_PE |
dc.title | Desarrollo de un filtro digital para señales foto pletismográficas obtenidas de una tarjeta de adquisición de datos en un entorno de laboratorio | es_PE |
dc.type | info:eu-repo/semantics/bachelorThesis | es_PE |
thesis.degree.level | Título Profesional | es_PE |
thesis.degree.grantor | Universidad Privada Antenor Orrego. Facultad de Ingeniería | es_PE |
thesis.degree.name | Ingeniero Electrónico | es_PE |
thesis.degree.discipline | Ingeniería Electrónica | es_PE |
dc.subject.ocde | https://purl.org/pe-repo/ocde/ford#2.11.00 | es_PE |
renati.advisor.orcid | https://orcid.org/0000-0002-6777-4479 | es_PE |
renati.author.dni | 76344364 | |
renati.author.dni | 74163148 | |
renati.advisor.dni | 44565065 | |
renati.type | https://purl.org/pe-repo/renati/type#tesis | es_PE |
renati.level | https://purl.org/pe-repo/renati/level#tituloProfesional | es_PE |
renati.discipline | 712026 | es_PE |
renati.juror | Cerna Sánchez, Elmer | |
renati.juror | Trujillo Silva, Marco | |
renati.juror | Ramos Rojas, Ovidio | |
dc.publisher.country | PE | es_PE |