dc.contributor.advisor | Villarreal Becerra, Einer Niels | |
dc.contributor.author | Cabanillas Martos, Maria del Carmen | |
dc.creator | Cabanillas Martos, Maria del Carmen | |
dc.date.accessioned | 2016-10-04T22:05:16Z | |
dc.date.available | 2016-10-04T22:05:16Z | |
dc.date.issued | 2016 | |
dc.identifier.uri | https://hdl.handle.net/20.500.12759/1906 | |
dc.description.abstract | Objetivo: Determinar la intensidad de luz emitida por unidades de fotopolimerización
utilizadas por cirujanos dentistas de la ciudad Cajamarca, 2015.
Materiales y métodos: El estudio prospectivo, transversal, descriptivo y
observacional, está constituido por 109 unidades de fotopolimerización utilizadas por
cirujanos dentistas de la ciudad Cajamarca en sus consultorios y/o clínicas privadas.
Primeramente se evaluó una serie de datos proporcionados por el cirujano dentista, que
se centra en el tipo de fuente de luz, marca y tiempo de uso de las unidades de
fotopolimerización; finalmente se determinó la intensidad de luz emitida por éstas con
ayuda de un radiómetro de marca Litex (Optilux Dentamerica iii ®). Para el análisis
estadístico los datos recogidos fueron procesados en el programa estadístico SPSS versión 21 (IBM).
Resultados: Los resultados obtenidos permitieron concluir que la gran mayoría de las
unidades de fotopolimerización presentan un nivel de intensidad de luz eficiente. En
relación al tipo de fuente de luz, marca comercial y tiempo de uso se ubican
mayormente en el nivel eficiente en emisión de luz.
Conclusión: El nivel de intensidad emitida por unidades de fotopolimerización de
cirujanos dentistas de la ciudad de Cajamarca es significativamente eficiente. | es_PE |
dc.description.abstract | Objective: To determine the intensity of light emitted by light curing units used by
dentists in Cajamarca city, 2015.
Materials and methods: The prospective, transversal, descriptive and observational
study consists of 109 curing units used by dentists in Cajamarca city in their offices
and / or private clinics. First, a series of data provided by dentists, which focuses on
the type of light source, make and time use of curing units was evaluated. Finally the
intensity of light emitted by curing units using a radiometer Litex mark (Dentamerica
® Optilux) was determined. For statistical analysis the data collected were processed
in SPSS version 21 (IBM).
Results: The results lead to the conclusion that the vast majority of light-curing units
had a level of light intensity efficiently. Regarding the type of light source, brand and
time use are located mostly in the efficient light emission level.
Conclution: The intensity of light emitted by light curing units used by dentists in
Cajamarca city is significantly efficient. | en_US |
dc.description.uri | Tesis | es_PE |
dc.format | application/pdf | es_PE |
dc.language.iso | spa | es_PE |
dc.publisher | Universidad Privada Antenor Orrego | es_PE |
dc.rights | info:eu-repo/semantics/openAccess | es_PE |
dc.rights.uri | https://creativecommons.org/licenses/by/4.0/ | es_PE |
dc.source | Universidad Privada Antenor Orrego | es_PE |
dc.source | Repositorio Institucional - UPAO | es_PE |
dc.subject | Intensidad de luz | es_PE |
dc.subject | Unidades de fotopolimerización | es_PE |
dc.title | Intensidad de luz emitida por unidades de fotopolimerización utilizadas por cirujanos dentistas de la ciudad de Cajamarca, 2015 | es_PE |
dc.type | info:eu-repo/semantics/bachelorThesis | es_PE |
thesis.degree.level | Título Profesional | es_PE |
thesis.degree.grantor | Universidad Privada Antenor Orrego. Facultad de Medicina Humana | es_PE |
thesis.degree.name | Cirujano Dentista | es_PE |
thesis.degree.discipline | Estomatología | es_PE |
dc.subject.ocde | https://purl.org/pe-repo/ocde/ford#3.02.14 | es_PE |
renati.type | https://purl.org/pe-repo/renati/type#tesis | es_PE |
renati.level | https://purl.org/pe-repo/renati/level#tituloProfesional | es_PE |
renati.discipline | 911016 | es_PE |
dc.publisher.country | PE | es_PE |